LC-MS may be performed on the Orbitrap only by special arrangement with the Facility, on a collaborative basis. Please contact the Facility Manager for details.
MS/MS may be performed by special arrangement with the Centre, usually on a collaborative basis. Please contact the Facility Manager for details. Different methods of tandem operation are available, in various ionisation modes:
Collision induced dissociation (CID), HCD (high energy CID) or PQD (pulsed-Q dissociation) on Orbitrap XL, with selection of precursor ions at low or high resolution and accurate mass of fragment ions (ESI, nESI or APCI)
High resolution instrument giving accurate mass measurement over the full mass range in electrospray/APCI
Commercial supplier: Thermo Scientific
High-resolution accurate-mass ESI, APCI, EScI, ASAP and APGC
Commercial Supplier: Waters
GC-MS is performed on the Waters GCT Premier using EI or CI, with ammonia (or methane), in either polarity. Accurate mass GC-MS can also be performed using the GCT Premier. The stationary phase of preference is DB5 (95% methyl and 5% phenyl). GC conditions depend on those specified in the analysis request.
Benchtop orthogonal acceleration time-of-flight (oa-TOF) mass spectrometer.
Commercial Supplier: Waters
MALDI is performed on Bruker ultrafleXtreme mass spectrometer. A typical procedure involves mixing solutions of sample and matrix, and pipetting 0.5-1 microlitre onto the target well of a sample plate. Inorganic salts are sometimes added to promote ionization. The sample spot is dried, allowing co-crystallisation of the mixture, then irradiated with a laser. Solvent-free preparations can also be used. The matrix is ablated from the plate while the sample is simultaneously desorbed and ionised, then accelerated into a flight tube (typically 20kV). The instrument may be run in any combination of positive or negative and linear or reflector modes. Specific details regarding conditions will accompany results.
Using the service -
more information about services available for different analytes (including polymer details).
Biochemicals - recommended sample
clean-up procedures.
Smartbeam-II™ laser enables ultra-high data acquisition speed in both MS and MS/MS
Commercial Supplier: Bruker
The TM-Sprayer™ Tissue MALDI Sample Preparation System is designed to provide the highest quality matrix deposition on the two-dimensional biological samples.
Commercial Supplier: HTX Imaging
ESI analyses are generally performed on the Orbitrap XL or Xevo G2-S using nanoelectrospray. Printouts will clearly state on the top whether nanoelectrospray or pneumatically-assisted electrospray has been used (see both methods below). The spectra achieved by both methods are similar; only the inlet and rate of sample consumption differ. For both spray rates, different source ("cone") voltages may be used to either promote molecular fragmentation ("harder" ionisation) or to minimise it and promote multiple charging ("softer" ionisation). Spectra are labelled appropriately when different source voltages give different results. Electrospray data acquired on the Orbtirap to m/z 2000 are generally also "accurate mass measurements" (5 ppm guaranteed; usually better than 3ppm).
Analyses are performed in positive or negative ionisation mode (one polarity per analysis). The Advion NanoMate inlet is used with a 96-well plate, corresponding transfer tips and 400-nozzle spray-chip. (With this inlet, each sample uses its own well, sample tip and nozzle, effectively eliminating the "memory" problem that is common with standard electrospray.) Solvents and additives used are listed on the results. Orbitrap XL conditions: see SOP document.
Electrospray: analyses are performed in positive or negative ionisation mode. Samples are loop injected into or infused in a stream of water / methanol (1:1 at 50µL/min.) using the customer-specified solvent for dissolution. Nebulisation is pneumatically assisted by a flow of nitrogen through a sheath around the capillary, a process sometimes referred to as "Ion spray". See Orbitrap SOP for more information on this instrument. Orbitrap XL conditions: see SOP document.
ESI sample requirements (wide range of analytes): especially suitable for polar or thermo-labile samples; ESI specialities are salts, non-covalently bound species, multiply charged complexes and very high MW (bio)molecules that will multiply charge. (In this laboratory, MALDI rather than ESI is generally used for high mass species.)
Accurate mass Electrospray (ESI) is the preferred laboratory method. Other techniques will only be used if ESI fails. If both a full-scan characterisation and accurate mass are required, both are done simultaneously. If only a single ion accurate mass measurement is required, it may be performed on the Orbitrap XL or on the Xevo G2S in positive or negative ionisation.
Orbitrap XL conditions: see SOP document. Standard operation uses nanoelectrospray on the Advion TriVersa NanoMate.
Advion TriVersa NanoMate chip-based nanoelectrospray infusion system
Commercial supplier: Advion
APCI in this laboratory is performed on the Orbitrap XL and on the Xevo G2S using the Atmospheric Solids Analysis Probe (ASAP) to insert samples into the APCI source. (Traditional flow injection APCI is also available but is rarely required). Samples are deposited onto a glass capillary as a solid or in solution. Additives such as ammonium acetate may also be used on the probe or in the ion source, to assist ionisation.
Atmospheric Solids Analysis Probe (ASAP) conditions:
Source Temp.: sample dependent. If the vapourizing temperature is unknown, the sample is introduced at ambient temperature (circa 40ºC) then the temperature is increased until the sample vaporizes.
Corona Discharge Current: 4 µA.
Accurate Mass APCI is also carried out using ASAP under the above conditions. For instrument calibration, please see Orbitrap SOP document (instrument is calibrated using ESI). Accurate mass is obtained on all ions in the spectrum simultaneously. APCI by direct infusion is available upon request.